• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558350
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2022.During the procedure, the balloon burst at 12 atm while being inflated inside the patient.It was reported that the patient had a very narrow stricture.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).
 
Additional Manufacturer Narrative
Block h6: device code a0402 captures the reportable event of balloon burst.Block h10: investigation result the returned cre fixed wire dilatation balloon was analyzed and a visual examination found that the balloon had two ruptures.The catheter of the device had no damages.Microscopic inspection found the balloon had two circumferential ruptures, located at 3 and 6 cm from the distal tip of the balloon.A functional inspection could not be performed due to the returned condition of the device.No other problems with the device were noted.With all the available information, boston scientific concludes the reported event of balloon burst was confirmed.The circumferential tear and rupture found in the balloon are likely to have occurred due to factors encountered during the procedure, such as excess pressure, interaction with other devices, or anatomical factors.It is possible that interaction with any kind of a sharp surface during or previous to the procedure could create friction on the balloon, causing physical damage.These factors and interactions could influence the problem found in the balloon.Therefore, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6), 2022.During the procedure, the balloon burst at 12 atm while being inflated inside the patient.It was reported that the patient had a very narrow stricture.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15808347
Report Number3005099803-2022-06703
Device Sequence Number5401630
Product Code KNQ
UDI-Device Identifier08714729202004
UDI-Public08714729202004
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K971320
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2024
Device Model NumberM00558350
Device Catalogue Number5835
Device Lot Number0029520010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-