| Catalog Number |
1C8581 |
| Medical Device Problem Codes |
Break (1069); Failure to Disconnect (2541)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the unspecified quantity of interlink system y-type solution sets would not disconnect to the patient¿s hickman (non-baxter).Several attempts were made to disconnect the set; however, the set would not disconnect and resulted in the tubing breaking off and stuck inside the hickman.This issue occurred during patient use of the device.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.Visual inspection was performed using the naked eye which observed the luer assembly was broken.Pressure testing was performed which revealed a leak in the affected area.Additional pull test was performed and no separation was observed.Due to the nature of the returned sample no additional testing could be performed.The reported condition was verified.The cause of the broken luer was due to the force applied by the user.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|