| Model Number |
CLH-SC |
| Medical Device Problem Code |
Electrical Power Problem (2925)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned and evaluated, and the customer¿s allegation that the lamp did not light was confirmed; no noise was generated, therefore that allegation was not confirmed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
| |
|
Event or Problem Description
|
|
The customer reported to olympus that when the halogen light source was turned on, an abnormal noise occurred, and the lamp did not light.The issue was found during preparation for use for a diagnostic procedure (colposcope), and the procedure was postponed because of the issue.There was no patient harm associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event was unable to be identified.It is likely that the lamp went out due to a disconnection of md-151 filament.The strange noise was not confirmed.Olympus will continue to monitor field performance for this device.
|
| |
|