CHRISTOPH MIETHKE GMBH & CO KG M.BLUE 5 SYS W/PED.CONTROL RESERVOIR; HYDROCHEPHALUS MANAGEMENT Back to Search Results
Model Number FX816T
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: external deposits on the reservoir.A permeability test has shown that the m.Blue has a blockage.The reservoir is permeable.Adjustment test: the m.Blue valve was tested and is not adjustable throughout the normal range.The brake functionality test has shown that the brake function operates as expected.However, the click force could not be measured due to the non-adjustability.Because the valve is not permeable, a computer controlled test is not possible.After dismantling of the valve, deposits were found in m.Blue for the sake of thoroughness and transparency, we would like to point out that an permeability test were already carried out after the revision at the hospital.Based on our investigation, we can determine a blockage and a non- adjustability of the m.Blue at the time of our investigation.The detected deposits can be named as the cause for the functional impairment.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a m.Blue (part # fx816t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt system was believed to be operated in underdrainage.The patient underwent a revision procedure performed on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 2 years, 2 months.Height: 84.5 centimeters (cm).Weight: 9.8 kilograms (kg).Gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M.BLUE 5 SYS W/PED.CONTROL RESERVOIR
Common Device Name
HYDROCHEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15809328
Report Number3004721439-2022-00344
Device Sequence Number12545978
Product Code JXG
UDI-Device Identifier04041906504526
UDI-Public4041906504526
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K192266
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/24/2025
Device Model NumberFX816T
Device Catalogue NumberFX816T
Device Lot Number20046433
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Initial Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age2 YR
Patient SexMale
Patient Weight10 KG
-
-