Model Number FV413T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation: visual inspection: a deformation of the outer housing of the progav valve was observed through the visual inspection.The measurement of the plane parallelism could confirm that with a value of -0,042mm the housing membrane is outside tolerance (0 ± 0.02 mm).Some deposits on the catheter is visible.Permeability test: a permeability test has shown that both valves are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the both valves are not operating within the acceptable tolerance in either position.Adjustment test: the progav was tested and is limited adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational, however the braking force cannot be measured due to the limited adjustability of the valve.Internal inspection : after dismantling of the valves, deposits were found in both valves.To make the proteins / deposits in the shunt system more visible, they were colored using a staining solution.Results : based on our examination results, we can determine an accelerated, slower outflow and limited adjustability on the shunt system.The cause of the slower/accelerated outflow and limited adjustability could be the detected deposits.Deposits caused by substances naturally present in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav shuntsystem (part # fv413t) was implanted during a procedure performed on an unknown date.According to the complainant, the shunt system was believed to be operated in overdrainage and difficult to adjust.The patient underwent a revision procedure.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 12 years.Height: 150 centimeters (cm).Weight: 40 kilograms (kg).Gender: male.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV WITH SHUNTASSISTANT 20
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15809931
Report Number3004721439-2022-00343
Device Sequence Number5615934
Product Code JXG
UDI-Device Identifier04041906132101
UDI-Public4041906132101
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2015
Device Model NumberFV413T
Device Catalogue NumberFV413T
Device Lot Number20011812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/17/2022
Initial Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/06/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age12 YR
Patient SexMale
Patient Weight40 KG
-
-