Model Number FREDX5026-PMA
Medical Device Problem Codes Stretched (1601); Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was implanted in the patient and is not available for return to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.
 
Event or Problem Description
It was reported that the fred stent was used to treat a 4.3mm distal x 5.0mm proximal left internal carotid artery (ica) aneurysm.The stent was implanted with proximal end twisted.The physician implanted a non-microvention stent in the closed segment followed by balloon angioplasty to ensure open inner lumen.Patient was stable during entire procedure and was doing well post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FRED X27
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key15810726
Report Number2032493-2022-00529
Device Sequence Number5472171
Product Code OUT
UDI-Device Identifier00842429114469
UDI-Public(01)00842429114469(11)211206(17)241130(10)0000131522
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P180027
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberFREDX5026-PMA
Device Catalogue NumberFREDX5026
Device Lot Number0000131522
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Initial Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-