• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES BD FACS¿ SAMPLE PREP ASSISTANT III; SEE H.10 Back to Search Results
Model Number 647205
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that while using the bd facs¿ sample prep assistant iii that the wash tower is overflowing and the lock is not functioning.The following information was provided by the initial reporter: wash tower overflowing and lock not functioning.
 
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Common device name: automated pipetting, diluting and specimen processing workstations for flow cytometric analysis.
 
Event or Problem Description
It was reported that while using the bd facs¿ sample prep assistant iii that the wash tower is overflowing and the lock is not functioning.The following information was provided by the initial reporter: wash tower overflowing and lock not functioning.
 
Additional Manufacturer Narrative
H.6 investigation summary pr6201282, pn: 647205, spaiii, sn: (b)(6), awareness: 09/30/2022 opened: 09/30/2022.¿ investigation summary: ¿ scope of issue: the scope of issue is limited to part: 647205 spaiii and serial number: (b)(6).¿ problem statement: customer reported: wash tower overflowing/door lock not working.¿ manufacturing defect trend: there are 0 qns related to the reported issue.Date range (date of incident to 12 months back) from (b)(6) 2021 to date (b)(6) 2022 (rolling 12 months).¿ complaint trend: there are 7 complaints related to the reported complaint.Date range (date of incident to 12 months back) from 30sep2021 to date 30sep2022 (rolling 12 months) complaint data attached.¿ investigation result / analysis: per fse¿s report: the leak was coming from the wash tower due to the waste line was backed up from blockage in the waste pump head assembly.The leak was waste, and fluids used for the instrument operations such as sheath, di water and lyse.There was no bodily contact or harm to the customer or fse caused by the leak however the leak was not contained within the instrument.Bleach was used to clean and disinfect the instrument and surrounding areas.The clog in the waste pump head was due to cores from the yellow vacutainer caps.The tubes used are bd vacutainer acd solution a #364606.The probe had 1800 piercings and was lot# f164970.The blockage was cleared after disassembling the head and flushing with di water.The pump was reassembled, and the couplings were replaced to avoid any further cores from going downstream past the filter.The filter was inspected.The problem was corrected and there was no further leaking or blockage.O second issue was the safety door lock not locking.This was due to bad sensors for the door as well as the solenoid sensors.After replacing the sensors, the problem was corrected.O tested and verified instrument performance.O no further issues.O sample preparation was not affected.O there was no delay in patient treatment due to any unexpected results.¿ service max review: review of related work order#: wo - (b)(4).Install date: 01sep2009.Defective part number: 364606 ¿ tube acd.Work order notes: o subject / reported: wash tower overflowing/door lock not working.O problem description: fluid leak/sensor failure.O cause: overflow caused by clogged pump and filters/sensor failure.O work performed: cleared blockage from pump/replaced sensor.O solution: cleared blockage from pump and fluidic lines/replaced sensor.O parts replaced: 364606 ¿ tube acd/ (sensor part number not reported).¿ returned sample analysis: the complaint sample was not requested to be returned and no photo(s) / picture(s) were provided.¿ manufacturing device history record (dhr) review: review of the dhr for serial number: (b)(6) and pn647205 was reviewed.The instrument met all the manufacturing specifications prior to release.¿ risk analysis: o risk management file part #100245ra, revision 03 was reviewed.O hazard(s) identified? ¿yes ¿no.¿ hazard id: 3.1.29.¿ hazard: environment biohazard.¿ cause: filter/components in fluid path get clogged.¿ harmful effects: exposure to biohazard.¿ residual severity: 5.¿ residual probability: 1.¿ residual risk index: 5.¿ root cause: based on the investigation result and fse¿s report the root cause was clogged pump and fluidic lines/sensor failure.¿ conclusion: based on the investigation results and the fse¿s report the complaint was confirmed for wash tower overflowing and door lock not working.No one was harmed or injured, and no medical diagnosis was performed due to any incorrect results.There was no impact to the patient health or safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD FACS¿ SAMPLE PREP ASSISTANT III
Common Device Name
SEE H.10
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer (Section G)
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
2350 qume drive
san jose CA 95131
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15811182
Report Number2916837-2022-00348
Device Sequence Number5578578
Product Code PER
UDI-Device Identifier00382906472052
UDI-Public00382906472052
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K131301
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number647205
Device Catalogue Number647205
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer12/30/2022
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date01/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1