Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer "picc was leaking, small pin hole near insertion site." additional information from customer: picc was leaking at insertion site, after line was d/c¿d, vat rn noticed a ¿pin hole¿ at the level of 0.Catheter was secured with a statlock.Line was removed and a second picc was inserted the next day.Patient is being treated for osteomyelitis.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regt6055 showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was confirmed.The product returned for evaluation was one 4fr s/l powerpicc solo catheter.Usage residues were observed throughout the sample and a needleless injection cap was attached to the luer adapter.A small, diagonally aligned split was observed between the molded joint and the 0cm depth marking.An attempt to infuse water through the sample using a 12ml syringe revealed the sample to be patent to infusion; however, a leak was observed at the split site.Microscopic inspection of the split revealed a granular fracture surface and inward curving edges.Multiple impressions were observed in the vicinity of the split and circumferentially opposite the split.The split features and abundant surface impressions were consistent with damage caused by contact between the catheter and a traumatic instrument, such as forceps or hemostats.The location of the damage suggested that catheter securement, maintenance and access techniques may have contributed.
 
Event or Problem Description
It was reported by the customer "picc was leaking, small pin hole near insertion site." additional information from customer: picc was leaking at insertion site, after line was d/c¿d vat rn noticed a ¿pin hole¿ at the level of 0.Catheter was secured with a statlock.Line was removed and a second picc inserted the next day.Patient is being treated for osteomyelitis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO2 CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15811475
Report Number3006260740-2022-05336
Device Sequence Number14841018
Product Code LJS
UDI-Device Identifier00801741155093
UDI-Public(01)00801741155093
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model NumberN/A
Device Catalogue Number1194108D4
Device Lot NumberREGT6055
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2022
Supplement Date Received by Manufacturer01/20/2023
Initial Report FDA Received Date11/16/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age53 YR
Patient SexMale
-
-