• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number VNMF4343C103TU
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/27/2021
Type of Reportable Event Death
Event or Problem Description
Valiant navions stent graft were implanted during the endovascular treatment of a thoracic aneurysm.  it was reported that on ct performed approximately 9 months post the index procedure an aortic stent graft was identified extending from the aortic arch to the distal descending thoracic aorta.Contrast was identified in the aneurysmal sac outside of the stent graft, consistent with an endoleak.The size of the aneurysmal sac of the descending thoracic aorta was relatively stable, measuring 84mm in maximum diameter.The endoleak is new compared with prior exam.  there was no  cause of endoleak provided.  the patient has died on an unknown date due to an unknown cause.No additional clinical sequelae was reported and the patient has expired.
 
Additional Manufacturer Narrative
Concomitant medical products: information references the main component of the system.Other relevant device(s) are: product id: vnmc4337c200tu, serial/lot #: (b)(4), ubd: 15-oct-2022, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Continued from h9 3005364322-02-19-2021-001-r.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Core lab review of ct imaging from approximately 3 days, 2 months and 4 months post index procedure observed no endoleak, stent ring enlargement or stent ring migration and were not evaluable for fracture.Core lab review of ct imaging from approximately 4 months post index procedure was not evaluable for endoleak, fracture or stent ring enlargement and no stent ring migration was observed.Core lab review of ct imaging from approximately 9 months post index procedure was not evaluable for endoleak or fracture and no stent ring enlargement or stent ring migration was observed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15812835
Report Number9612164-2022-04281
Device Sequence Number5837975
Product Code MIH
UDI-Device Identifier00763000100759
UDI-Public00763000100759
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2021
Device Model NumberVNMF4343C103TU
Device Catalogue NumberVNMF4343C103TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer05/03/2023
05/28/2024
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date05/04/2023
06/26/2024
Date Device Manufactured02/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
SEE H10
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Outcome Attributed to Adverse Event Death;
Patient Age67 YR
Patient SexMale
Patient Weight64 KG
-
-