BIOMET 3I OSSEOTITE¿ TAPERED CERTAIN¿ IMPLANT 6 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number XIFNT610
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Event or Problem Description
It is reported that the patient requested that the clinician remove the implant.Tooth site #30 (universal).
 
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Patient identifier is not provided / unknown.Patient age is not provided / unknown.Patient weight is not provided / unknown.Date of event is not provided / unknown.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.The following sections are being reported: h2: if follow-up, what type.H3: device evaluated by manufacturer.H6: type of investigation.H6: investigation findings.H6: investigation conclusions.H10: additional narratives/data.The product was returned for investigation.The reported event is non-verifiable.Based on the evaluation, the device malfunction has not occurred.Additionally, there is no existing nonconformance/capa/hhe/d/ie/product holds against the reported devices that could cause or contribute to the reported event.Zimmer biomet quality management system (qms) has controls in place to ensure the distribution of conforming product.Therefore, based on the available information, the products were likely within specifications and likely conforming when they left zimmer biomet.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number for similar event and no other complaint was identified.A definitive root cause could not be identified.No further investigation and no immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimvie will continue to monitor for trends.
 
Event or Problem Description
No additional or corrected information to report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE¿ TAPERED CERTAIN¿ IMPLANT 6 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15813479
Report Number0001038806-2022-01777
Device Sequence Number7133163
Product Code DZE
UDI-Device Identifier00844868022053
UDI-Public(01)00844868022053(17)251130(10)2020111040
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K130949
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date11/30/2025
Device Model NumberXIFNT610
Device Catalogue NumberXIFNT610
Device Lot Number2020111040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer03/09/2023
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date03/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexPrefer Not To Disclose
-
-