| Model Number |
G38495 |
| Medical Device Problem Codes |
Off-Label Use (1494); Difficult or Delayed Activation (2577)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Per email from dm: product malfunction after the stent was deployed in the left brachial artery, the deployment mechanism would come off the wire.The device got stuck and (b)(6) had to take out the wire and deployment mechanism together.(b)(6) inserted a 180cm guidewire to complete the case.There were no adverse events and the case was completed without further complications.Patient outcome: did any unintended section of the device remain inside the patient¿s body? no.Was the patient hospitalized or was there prolonged hospitalization? no.Did the patient require any additional procedures due to this occurrence? no.Did the product cause or contribute to the need for additional procedures? no.Has the complainant reported any adverse effects on the patient due to this occurrence? no.Has the complainant reported that the product caused or contributed to the adverse effects? n/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # p100022/s014.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation the zisv6-35-125-8-40-ptx device of lot number c1934938 involved in this complaint was returned for evaluation, without the original packaging.With the information provided, a physical examination and document-based investigation was conducted.Lab evaluation the device related to this occurrence underwent a laboratory evaluation on the 28th feb 2022.On evaluation of the device, the following was noted: visual inspection: red safety tab is depressed on return.0.035" wireguide still inserted in the device on return.The wire guide returned in the device was measured using the vernier callipers and was found to be 0.035¿.This is the required wire guide size for use with this device as per the ifu and product label.No stent returned.Functional inspection: wireguide removed with no issue.Device flushed with no issue.Document review prior to distribution all zilver ptx drug-eluting peripheral stent devices are subjected to a visual inspection and functional inspection to ensure device integrity.A review of the manufacturing records did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records confirms the failure mode has not previously occurred for this work order.It should be noted that the instructions for use (ifu0118) states the following: ¿the zilver ptx drug eluting peripheral stent is intended for use in the treatment of symptomatic vascular disease of the above the knee femoro-popliteal arteries.¿ there is evidence to suggest the user did not follow the ifu.From the information provided, it is known that the device was placed during a left arm fistula angioplasty and stent placement and the target location was the left subclavian vein which is not the intended use of this device.Image review: an image was not returned for evaluation.Root cause review: a definitive root cause of off label use can be concluded.From the information provided, it is known that the device was placed during a left arm fistula angioplasty and stent placement and the target location was the left subclavian vein which is not the intended use of this device.These devices have not been designed, sized, tested or risk assessed for use in this anatomy.The ifu states ¿the zilver ptx drug eluting peripheral stent is intended for use in the treatment of symptomatic vascular disease of the above the knee femoro-popliteal arteries¿.It should be noted that, as the device was used outside of their validated state and/or against the instructions provided in the ifu, it is not possible to predict how the devices will perform or function.As per d00213689, rev006, section 5.3 table 1 - off label use complaints are considered to be unforeseen misuse.It is unknown how the device will function outside of its intended use.Summary: the complaint is confirmed based on customer testimony.According to the initial reporter, the patient did not suffer any adverse effects due to this occurrence.The doctor took out the wire and deployment mechanism together and completed the the procedure with no further complications.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Event or Problem Description
|
|
A supplemental report is being submitted due to the completion of the lab evaluation on 28-feb-2023 and the completion of the investigation on 14-mar-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Pma/510(k) # p100022/s014.Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
A supplemental report is being submitted due to: off label use confirmed by medical affairs and engineering on 23-jan-2023: left subclavian vein is off label.
|
| |
|