BOSTON SCIENTIFIC CORPORATION RELIANCE 4-FRONT; IMPLANTABLE LEAD Back to Search Results
Model Number 0663
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Under-Sensing (1661)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 11/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited undersensing on the atrial channel.This resulted in inappropriate anti-tachycardia pacing (atp) and multiple electric shocks due to an atrial driven arrhythmia.Boston scientific technical services (ts) was consulted and discussed programming options as well as medication options.No additional adverse patient effects were reported.At this time, this device system remains in service.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) exhibited undersensing on the ventricular channel.This resulted in inappropriate anti-tachycardia pacing (atp) and multiple electric shocks due to an atrial driven arrhythmia.Boston scientific technical services (ts) was consulted and discussed programming options as well as medication options.No additional adverse patient effects were reported.At this time, this device system remains in service.
 
Additional Manufacturer Narrative
Corrected fields: b5 describe event or problem model and serial number of the reported lead have been corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIANCE 4-FRONT
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15814679
Report Number2124215-2022-47658
Device Sequence Number20183986
Product Code LWS
UDI-Device Identifier00802526592805
UDI-Public00802526592805
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S145
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/15/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/22/2021
Device Model Number0663
Device Catalogue Number0663
Device Lot Number108744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/08/2022
Supplement Date Received by Manufacturer12/13/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
-
-