| Catalog Number |
00588005012 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an initial right total knee arthroplasty.Subsequently, between six to eighteen months post-implantation, the rotating hinge post backed out.The patient underwent revision surgery to replace the articular surface and hinge post.It was reported that no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Implantation date occurred in 2021 though exact date is unknown.Concomitant medical products: unknown rotating hinge knee femoral component: catalog#ni, lot#ni; unknown rotating hinge knee tibial tray: catalog#: ni, lot#: ni.Customer has indicated that the product will not be returned to zimmer biomet for evaluation per hospital policy.The investigation is in progress.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.D10: medical product: right size f cemented option femoral component: catalog#00588001602, lot#64695386; 15mm diameter 30mm length straight stem extension combined length 75mm: catalog#00598801215, lot#64560563; size 4 precoat cemented tibial component: catalog#00588000400, lot#65069785; all poly patella cemented 35 mm diameter: catalog#42540000035, lot#64995552; unknown stryker cement: catalog#ni, lot#ni, qty#3; unknown stryker cement restrictor: catalog#ni, lot#ni, qty#(b)(4).Additional information provided has prompted a review of the previously reported investigation results.The investigation is in progress.Upon completion of the investigation, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an initial right total knee arthroplasty.Subsequently, seven (7) months post-implantation, the patient began to experience instability and underwent revision surgery where it was determined that the femoral hinge post extension disassociated from the articular surface.The polyethylene bearing was exchanged for a larger size and a new hinge post extension was locked into place.All available information has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: stability exam revealed instability to valgus stress with mechanical dissociation of the hinge apparatus; hinge post had dissociated from the femoral portion of the hinge and the tibial poly bearing was no longer stabilized.12mm poly was removed and replaced with 14mm poly.New hinge post was tightened manually then locked into place with torque wrench.No intraoperative complications were found.Investigation results concluded that the reported event is attributed to failure of the user to follow instructions.The reported products were reviewed for compatibility and it was determined that the implanted articular surface and femoral component are not compatible.Detailed instructions for properly matching components are provided in the ifu.The device instructions for use states the following: to properly match the components, the femoral and tibial component size must both be reflected on the articular surface label.Mismatching may result in poor surface contact and cause pain, greater wear, implant instability or otherwise reduce implant life.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.Reported event was confirmed by review of provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information at time of this report.
|
| |
|