• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-FV-14-T; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-FV-14-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-FV-14-T
Medical Device Problem Codes Deflation Problem (1149); Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformities noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported right side "rupture" of saline device.Device has been explanted and replaced with implant.
 
Event or Problem Description
Healthcare professional reported right side "rupture" of saline device.Device has been explanted and replaced with implant.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed, opening device on the edge of the insert to shell bond side assessed as unidentified (tear) opening (shell thickness within specification) and one opening in stable base assessed as surgical damage.Additional observations: ¿ yellow particles in device inner surface are observed.¿ broken of suture tab assessed as unidentified (tear) opening.No further actions are required as the device was implanted.Additional, changed, and/or corrected data: d9, h3, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-FV-14-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15815248
Report Number9617229-2022-20407
Device Sequence Number20172072
Product Code LCJ
UDI-Device Identifier10888628043800
UDI-Public10888628043800
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K102806
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/17/2025
Device Model Number133S-FV-14-T
Device Catalogue Number133S-FV-14-T
Device Lot Number3421682
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date12/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-