|
|
| Model Number |
Z497G |
| Medical Device Problem Codes |
Break (1069); Material Frayed (1262)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation procedure on (b)(6) 2022 and suture was used.During use on the patient, when trying to use the sutures and put it in place, the sutures were fraying then breaking.A different box with a different lot number was used to complete the procedure.There were no adverse patient consequences.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.(b)(4).A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Related events captured via 2210968-2022-09531, 2210968-2022-09532, 2210968-2022-09533, 2210968-2022-09535, 2210968-2022-09536.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Date sent to the fda: 12/21/2022.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional test evaluation were conducted on the returned devices.Visual analysis of the returned sample determined that it was received one open box with eight unopened samples pertain to the product code z497.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no issue related to breakage suture or anomalies were observed during evaluation.In addition, a functional test was performed using an instron equipment and the tensile force was above the minimum requirements.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2022-09531, 2210968-2022-09532, 2210968-2022-09533, 2210968-2022-09534, 2210968-2022-09535, 2210968-2022-09536.
|
| |
|
Search Alerts/Recalls
|
|
|