|
It was reported that, during an ankle reconstruction, the biosure screw broke inside the patient.The fracture pieces were removed from the patient by tweezers.The surgery was resumed after a non-significant surgical delay using a back-up device.No further complications were reported.
|
|
It was reported that, during an ankle reconstruction, the biosure screw broke inside the patient.The broken pieces were removed from the patient by tweezers.The surgery was resumed after a non-significant surgical delay using a back-up device.No further complications were reported.
|
|
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.A visual inspection of the returned device found that it is not in its original packaging.The returned device has the distal portion fractured away and broken into several pieces.The device has debris on it.A review of the customer provided image finds the complaint device, fractured into several pieces.Based on the condition of the product material found during visual inspection, additional material testing is not required.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.Since there has not been any harm to this patient because of the breakage screw breakage or the non-significant surgical delay, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|