• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STLESS STEEL SIL 4X18IN 5 S/A CCS; SUTURE, NONABSORBABLE, STEEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STLESS STEEL SIL 4X18IN 5 S/A CCS; SUTURE, NONABSORBABLE, STEEL Back to Search Results
Model Number M653G
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: was there any change in the patient¿s post-operative care due to the prolonged procedure? no.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? no.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Events reported via: 2210968-2022-09586.
 
Event or Problem Description
It was reported that a patient underwent a sternotomy procedure on (b)(6) 2022 and suture was used.During the procedure, in a trauma surgery opens aciflex caliber 5 and it is observed that the diameter was smaller than the size that is real.Very thin suture gauge was observed.Another suture is opened and the same is observed.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The following information was obtained: the sample is en route.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 4/25/2023 additional information: d9, h3 a device has been received for product code m653g lot ap9632, however clarification is requested whether the product belongs to this complaint.Our failure analysis lab received the additional products with reference to this complaint, the following product was received: - product code: m653g - product received: ap9632 this complaint is for product code m653g, lot an8487 1.Could you please clarify if additional device belongs to this complaint? 1a.If yes, did a device malfunction occur during the same procedure for both lots ap9632 & an8487? 2.If not, could you please clarify if the additional received product belongs to a different complaint? if so, can you please provide the complaint number.3.If the device was not reported before, please provide more details of device malfunction.4.If the product doesn¿t belong to any complaint, please let us know so we can discard it or advise if the product is required to be returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2022-09587, 2210968-2022-09586.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STLESS STEEL SIL 4X18IN 5 S/A CCS
Common Device Name
SUTURE, NONABSORBABLE, STEEL
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-BRAZIL
rodovia presidente dutra
km 154
sao paolo
BR  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15818447
Report Number2210968-2022-09587
Device Sequence Number7101823
Product Code GAQ
UDI-Device Identifier10705031045477
UDI-Public10705031045477
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model NumberM653G
Device Catalogue NumberM653G
Device Lot NumberAN8487
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2022
Supplement Date Received by Manufacturer02/06/2023
04/11/2023
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date02/13/2023
04/25/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-