MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM 3D; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number QC-7-30-3D
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report regarding two axium coil detachment failures.It was reported that the coils could not detach in the aneurysm.Then it was thrown out and instead used other new axium coil.Then completed this case.For both axium coils the physician attempted to detach the coil with the instant detacher three to four times.For one axium coil (lot#223258185) the physician attempted three to four times with a second instant detacher.There was a manual attempt at detachment via hypotube and was attempted three to four times.The second axium coil (lot# 224405948) detachment method was unknown.There was no issue with the instant detacher for either coil noted.The reported device and any accessory devices prepared as indicated in the instructions for use (ifu).No patient symptoms or further complications were reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that prior to coil non-detachment, there were no issues encountered.There was no damage to the pushwire.
 
Additional Manufacturer Narrative
H3.Product analysis #(b)(4):equipment used- video inspection system (m-81805), ruler (m-83360), camera (panasonic lumix dmc-zs5), in-house instant detacher (model: id-1-5 lot: 9443624) as found condition- one axium coil and one unknown non-medtronic coil were returned for analysis within a shipping box; within individual sealed plastic biohazard pouches; within medtronic dispenser coils and within medtronic introducer sheathes.The instant detachers used during the event were not returned for analysis.Visual inspection/damage location details: (axium coil) no damages or irregularities were found with the introducer sheath.The actuator interface was found securely attached to the coupler tube and the break indicator was found intact.There were no evidence of detachment attempts at these locations using an instant detacher or by manual method.No damages were found with the axium pusher.The coin was found still against the lumen stop.The shield coil was found undamaged.The implant coil was found still attached to the pusher.No damages or irregularities were found with the implant coil.Testing/analysis ¿ (axium coil) an in-house instant detacher was used for detachment testing.No difficulty was experienced inserting the pusher into the instant detacher, and the load indicator was not visible when the pusher was fully seated in the instant detacher cap, which is normal.The implant coil was successfully detached on the first attempt without any difficulty.Conclusion - based on the customer report and device analysis, the customer report of "non-detachment" could not be confirmed and the cause could not be determined.The failure could not be replicated as the device detached with no issues on the first attempt using an in-house instant detacher.No damages were found with the device that would contribute towards the failure.Customer reported attempting to detach the device using multiple instant detachers and by breaking the hypotube multiple times.However, no such breaks were found with the returned axium coil.The instant detachers used in the event were not returned for analysis.Therefore, any contribution of the instant detachers towards the non-detachment and conformance to specification could not be assessed.H6.Coding updated based on analysis results.Associated with mdr #: 2029214-2022-01955 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM 3D
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key15819299
Report Number2029214-2022-01954
Device Sequence Number7098688
Product Code KRD
UDI-Device Identifier00847536037878
UDI-Public00847536037878
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/05/2024
Device Model NumberQC-7-30-3D
Device Catalogue NumberQC-7-30-3D
Device Lot Number223258185
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2022
Supplement Date Received by Manufacturer11/17/2022
12/18/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date12/09/2022
01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-