• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ VIO 8X18IN 2-0 S/A CT CR; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ VIO 8X18IN 2-0 S/A CT CR; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number VCP751D
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2022 and suture was used.The needle broke when sewing in the surgery.Changed to another five to continue the surgery, but the same problem happened.Changed to another one to complete the surgery.The broken needle fell into the patient, but was retrieved shortly.The procedure was completed and there were no adverse patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested and the following was received: did the needle fall into the patient? yes.If yes, was the needle piece(s) retrieved during the same procedure? yes.Were x-rays taken to locate the needle piece(s)? unk.What measures were taken to retrieve the broken piece? unk.Was there any additional tissue damage as a result of searching for the needle piece?unk.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.(b)(4).A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.Trade name - irgacare®, active ingredient(s) ¿ triclosan, dosage form ¿ suture/solid/parenteral, strength ¿ = 275 g/m.Related events captured via 2210968-2022-09592, 2210968-2022-09594, 2210968-2022-09595, 2210968-2022-09596, 2210968-2022-09597.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: d9.H3 evaluation: the product was returned for evaluation.Visual inspection was conducted on the returned devices.Visual analysis of the returned sample revealed that it was received one open box with six unopened samples that pertain to product code.In order to evaluate the condition of the returned samples, the packets were opened, and the swage and attachment area was noted to be as expected.The tip and body of the needles were examined under magnification and no defects or needle breakage were found.In addition, the samples were tested by resistance test to needle and met the requirements.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch, and no non-conformances were identified.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 8X18IN 2-0 S/A CT CR
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
ciudad juarez 32604
MX   32604
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15819773
Report Number2210968-2022-09593
Device Sequence Number11541680
Product Code GAM
UDI-Device Identifier10705031055292
UDI-Public10705031055292
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberVCP751D
Device Catalogue NumberVCP751D
Device Lot NumberRBMPJQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-