Patient identifier: requested, not provided, age & date of birth: requested, not provided, patient sex: requested, not provided, weight: requested, not provided , ethnicity: requested, not provided, race: requested, not provided , implanted date: device was not implanted, explanted date: device was not explanted, occupation: lead tech.Since the actual sample was not returned to ashitaka factory, investigation of it was unable to be performed.The provided photo was confirmed.It was found that the outer layer of guidewire was peeled off.Review of the manufacturing record and the shipping inspection record of the involved product code/lot number combination confirmed that there were not any indications of anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot number combination.Based on the investigation result, no anomaly was found in the manufacturing record and the shipping inspection record.However, since the actual product was not returned and investigation could not be performed, it was not possible to clarify the cause of occurrence.Ashitaka factory assures the quality of this product by performing following controls.The guidewire is passed through the dedicated tool on 100% basis to confirm that there is no peeling of the outer layer or adhesion of any foreign substances on the surface of guidewire.Before the packaging process, 100% visual inspection is performed to confirm that there is no exposure of the core wire or anomaly in its appearance.Instructions for use (ifu) of this product includes the following warnings."if any resistance is felt or if the tip's behavior and/or location seems improper, stop manipulating the glide wire m and/or the catheter and determine the cause by fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the guide wire's tip, damage to the catheter, or damage to the vessel." "do not manipulate or withdraw the glidewire through a metal entry needle or a metal dilator.Manipulation and/or withdrawal through a metal entry needle or a metal dilator may result in destruction and/or separation of the outer polyurethane coating requiring retrieval.A plastic entry needle is recommended when using this wire for initial placement." terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
|