| Catalog Number |
BMK6F105MBER130 |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The patient was undergoing a medical procedure using a benchmark 6f 071 delivery catheter (benchmark).During the procedure, after a few minutes of use, the physician noticed leaking from a hole at the proximal end of the benchmark.Therefore; the benchmark was removed.The procedure was completed with a new benchmark.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned benchmark confirmed a leak beneath the stain relief.Evaluation revealed a fracture underneath the strain relief.If the benchmark is manipulated at extreme angles during use, damage such as a kink and subsequent fracture may occur.This damage likely contributed to the reported complaint during the procedure.Further evaluation revealed a distal shaft ovalization.This damage was likely incidental to the complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|