• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZA9003
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after an intraocular lens (iol) was fully inserted, the doctor noticed a broken haptic.The lens was removed and another johnson & johnson lens backup lens with the same model and diopter was used as the replacement.There were no medical or surgical interventions reported.The patient was reported to be doing fine and no patient injury reported.No further information was provided.
 
Additional Manufacturer Narrative
Age, weight, ethnicity: information unknown/not provided.Date of event: the exact date is unknown, the best estimate is prior to (b)(6) 2022.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.The intraocular lens (iols) has not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt has been made to obtain the missing information; however, to date, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: additional information received with the returned device had mentioned "16-0 inserted".Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: dec 21, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside of the original lens case.Patient stickers and the original folding carton were received as well.Additional information on the folding carton was also received.Visual inspection under magnification revealed that the complaint lens was received crushed inside of the lens insert and with a haptic detached.The lens was cleaned and, lens damage consistent with a lens that was crushed inside of the lens insert was observed.No further issues were observed.The complaint issue could not be confirmed.However, the observed issue haptic detached is similar to the complaint issue haptic damaged and could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Corrected data: in review, it was notice that in the section h10 of the initial mdr report, it was inadvertently indicated that "the exact date of event is unknown" which is incorrect as, the date of event (10/20/2022) had already been captured in the section b3" of the initial mdr report.This report is to correct that statement.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15820490
Report Number3012236936-2022-02867
Device Sequence Number11535599
Product Code HQL
UDI-Device Identifier05050474528949
UDI-Public(01)05050474528949(17)270802
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZA9003
Device Catalogue NumberZA90030165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer01/03/2023
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-