• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number DIU150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138); Visual Disturbances (2140)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from the patient's operative eye due to patient complained of blurred vision, anisometropia, and glare at night.Patient also experienced diplopia, therefore, patient was unhappy with vision.Visual acuity pre-operative (op): 20/30- and visual acuity post-op: 20/30-.There was no vitrectomy or sutures required, however, the incision was enlarged.Another johnson & johnson lens, same model diu150 21.5 diopter was implanted as a replacement.Through follow-up, it was learned it was the patient's right eye.The lens exchange was successful and the patient is satisfied.No other information was provided.
 
Additional Manufacturer Narrative
Ethnicity: unknown/asked but unavailable.Health effect clinical code: 2138 unexpected postop refraction, 2138 diplopia.Device evaluation: the device was not returned at the manufacturing site; therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.A search of complaints related to this production order (po) was performed.The search of complaints revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.An attempt was made to obtain missing information; however, the account did not provide the information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: dec 19, 2022.Device evaluation: the suspect lens was received inside a specimen cup.Visual inspection under magnification revealed that the lens was received cut in half and coated in viscoelastic residue.The lens was cleaned and inspected under magnification and lens damage and scratches were observed on the lens optic and haptics.The complaint issue "explant", "blurry", "diplopia", "glare" and "unexpected postop refraction" were not identified during product evaluation.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15820504
Report Number3012236936-2022-02838
Device Sequence Number11515810
Product Code MJP
UDI-Device Identifier05050474746619
UDI-Public(01)05050474746619(17)240202
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/02/2024
Device Model NumberDIU150
Device Catalogue NumberDIU150U235
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer12/23/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight84 KG
Patient RaceWhite
-
-