MEDTRONIC HEART VALVES DIVISION EVOLUT PRO TRANSCATHETER AORTIC VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number EVOLUTPRO-29
Device Problems Device Dislodged or Dislocated (2923); Activation Problem (4042)
Patient Problems Cardiac Arrest (1762); Low Cardiac Output (2501)
Event Date 11/12/2022
Event Type  Injury  
Manufacturer Narrative
Product analysis: the device remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Conduction disturbances are known potential adverse effects associated with any cardiac or thoracic procedure (open or catheter-based) and can be resolved with medical treatment or the implant of a permanent pacemaker (with the risk-benefit ratio in favor of implant of the percutaneous aortic valve).A conduction disturbance does not indicate a device malfunction or potential manufacturing issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that prior to implant of this transcatheter bioprosthetic valve, a balloon valvuloplasty with a 18 x 40 balloon was performed.Of note, the patient had aortic stenosis, aortic valve area (ava) 0.7, mean gradient 45, and little calcium observed on computed tomography (ct).The valve was positioned at a depth of 3 millimeter (mm).Incomplete expansion of the valve and low diastolic pressure was observed.It was decided to reposition the valve but during the attempt to reposition the valve, the valve was released and dislodged.The valve was moved to the ascending aorta and post dilatations were performed with a 20 x 40 balloon."significant resistance" was noted and a second valve was implanted.The patient recovered from three cardiac arrest and left the ward with inotropic support.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Updated b5.Corrected h6 - device conclusion code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received which reported that the valve was attempted to be recaptured, however was actually released.The reason for the recapture was incomplete valve expansion.The "significant resistance" reported was during use with the delivery catheter system (dcs).Following the implant of the valves, the gradient had reduced to 5 mmhg.
 
Manufacturer Narrative
Additional information was received that the treatment for the three cardiac arrests was dilating and repositioning the valves.The "significant resistance" was due to the amount of calcium in the leaflets of the native valve.The patient was discharged.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Per the device instructions for use (ifu), in the event that valve function or sealing is impaired due to excessive calcification or incomplete expansion, a post-implant balloon dilation of the bioprosthesis may improve valve function and sealing.To ensure patient safety, valve size and patient anatomy must be considered when selecting the size of the balloon used for dilatation.The balloon size chosen for dilatation should not exceed the diameter of the native aortic annulus.Refer to the specific balloon catheter manufacturer's labeling for proper instruction on the use of balloon catheter devices.With the limited information available, a conclusive root cause for the under expansion cannot be determined.It was also reported that the patient had aortic stenosis, aortic valve area (ava) 0.7, and a mean gradient of 45 millimeters of mercury (mmhg).Potential factors that can influence a dislodged valve include tension applied on the delivery catheter system (dcs) during positioning, calcification levels in the native vessel, compliance of the aorta and native vessels, and a number of others.In this case, it was reported that the valve was attempted to be recaptured, however was actually released and subsequently dislodged.With the limited information available, however, a root cause for the dislodgement could not be conclusively determined.Cardiac arrest is a known potential adverse effect per device ifu.In this case, a conclusive cause could not be determined from the limited information available, but the dislodgement was a likely contributing cause.Low cardiac output was identified as an event of hemodynamic instability and is a known potential adverse event per the device ifu.Low cardiac output (hemodynamic issue) can be related to valve related (degeneration, thrombus, calcification, etc.) or non-valve related factors (left ventricular outflow tract (lvot) obstruction, patient pressures, left ventricle dysfunction, etc.).A conclusive cause could not be determined from the limited information available.This event does not indicate device misuse or malfunction.Updated: h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT PRO TRANSCATHETER AORTIC VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15820639
MDR Text Key303898372
Report Number2025587-2022-03222
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/28/2023
Device Model NumberEVOLUTPRO-29
Device Catalogue NumberEVOLUTPRO-29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2022
Initial Date FDA Received11/17/2022
Supplement Dates Manufacturer Received11/23/2022
12/05/2022
03/24/2023
Supplement Dates FDA Received12/03/2022
12/12/2022
04/03/2023
Date Device Manufactured06/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight60 KG
-
-