| Model Number |
C8404 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [facility] incoming inspection po#(b)(6).We would like to inform you that some of the products didn't pass our receiving inspection.Please find attached spreadsheet.C8404 lot #1453797 (1 ea.) - particulate: hair.Products available for return.Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair likely originated from an operator during the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: [distributor] incoming inspection po#(b)(4).We would like to inform you that some of the products didn't pass our receiving inspection.Please find attached spreadsheet.C8404 lot #1453797 (1 ea.) - particulate: hair products available for return.Patient status: na (no patient involvement).Type of intervention: na (incoming inspection).
|
| |
|
Search Alerts/Recalls
|