Model Number IPN920396
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported the catheter was leaking at the extension line.The doctor inserted a primed catheter in the patient and a leak was displayed during blood draw.The cathter was replaced with a new picc set.No patient harm was reported.The patient's condition is reported as fine.
 
Event or Problem Description
It was reported the catheter was leaking at the extension line.The doctor inserted a primed catheter in the patient and a leak was displayed during blood draw.The cathter was replaced with a new picc set.No patient harm was reported.The patient's condition is reported as fine.
 
Additional Manufacturer Narrative
(b)(4).The actual device was not returned; however, the customer provided three photos for analysis.The first photo displays the lidstock and the second two photos display a 2-l cvc catheter in a tray.No damage is observed in the photo.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and one finding was identified.A non-conformance was initiated for 13c22a0158 and 13c21k2368 for the issue of component missing.This non-conformance is not related to the customer complaint.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.Do not secure, staple and/or suture directly to outside diameter of catheter body or extension lines to reduce risk of cutting or damaging the catheter or impeding catheter flow.Secure only at indicated stabilization locations." the customer report of an extension line leak was not able to be confirmed by visual inspection of the customer supplied photos.Full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PICC SET: 2-LUMEN 5 FR X 19-1/2" (50 CM)
Common Device Name
CATHETER, INTRAVASCULAR, THER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15821382
Report Number9680794-2022-00725
Device Sequence Number20372717
Product Code LJS
UDI-Device Identifier10801902194388
UDI-Public10801902194388
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K073451
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberIPN920396
Device Catalogue NumberPR-05052-LW
Device Lot Number14F22B0056
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2022
Supplement Date Received by Manufacturer12/14/2022
Initial Report FDA Received Date11/17/2022
Supplement Report FDA Received Date12/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED.; NOT REPORTED.
Patient SexUnknown
-
-