STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0° X3 INSERT 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 623-00-36F
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 06/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.The following devices were also listed in this report: cat# 502-03-60f; primary tritanium hemi cluster hole cup 60mm; lot#: mmjt6y.Cat# 6260-9-136; v40 cocr lfit head 36mm/0; lot#: mmn7pt.Cat# 6721-0937; size 9 accolade ii 127 deg; lot#: unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
A follow-up questionnaire was completed by the subject where they answered "no" for satisfaction with the results of the study hip replacement.The subject mentioned that they are not satisfied with their hip replacement and have daily pain in the left hip.Furthermore, the subject explained that the they have more pain now compared to before replacement.No additional details can be provided as the questionnaires were administered via mail and no clinical notes were obtained.
 
Event or Problem Description
Left hip.A follow-up questionnaire was completed by the subject where they answered "no" for satisfaction with the results of the study hip replacement.The subject mentioned that they are not satisfied with their hip replacement and have daily pain in the left hip.Furthermore, the subject explained that the they have more pain now compared to before replacement.No additional details can be provided as the questionnaires were administered via mail and no clinical notes were obtained.
 
Additional Manufacturer Narrative
An event regarding pain involving a trident liner was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.The following devices were also listed in this report: cat# 502-03-60f; primary tritanium hemi cluster hole cup 60mm; lot# mmjt6y.Cat# 6260-9-136; v40 cocr lfit head 36mm/0; lot# mmn7pt.Cat# 6721-0937; size 9 accolade ii 127 deg; lot# unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0° X3 INSERT 36MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15821963
Report Number0002249697-2022-01662
Device Sequence Number17681037
Product Code LPH
UDI-Device Identifier07613327039535
UDI-Public07613327039535
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2018
Device Model Number623-00-36F
Device Catalogue Number623-00-36F
Device Lot NumberMMJED9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer02/05/2024
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date02/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/09/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age79 YR
Patient SexMale
Patient Weight110 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-