H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.Images were provided.All images show delamination of the guidewire coating and decoiled tip of the guidewire.Decoiling of the guidewire can occur during heavy blockage of the catheter when the guidewire start to move with the inner helix.The break of the guidewire can be result of blockage or aspiration of very thick thrombotic material that results in catheter overheating, moving together with the guidewire, guidewire overheating and break.Based on the images it can be confirmed that guidewire was broken.A clear root cause could not be identified but a broken guidewire represents a known inherent risk.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 05/2024), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|