• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRGAS THERAPEUTICS LLC ISPAN PERFLUOROPROPANE (C3F8) GAS; INTRAOCULAR GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRGAS THERAPEUTICS LLC ISPAN PERFLUOROPROPANE (C3F8) GAS; INTRAOCULAR GAS Back to Search Results
Catalog Number 8065797105
Medical Device Problem Codes Gas Output Problem (1266); Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report originally filed under mfg report num 1610287-2022-00028.All future reports will be submitted under mfg report num reflected in this report.The c3f8 gas tank was not returned for evaluation.A review of the batch production record could not be performed because the lot number was not reported.The expiration date could not be determined.A review of the complaint records could not be performed because the lot number was not provided.A review of confirmed complaints for leaking cylinders of this type or for c3f8 not lasting as long as expected.Both were for cylinders that leaked at the valve to cylinder connection.The root cause of the reported event cannot be determined.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that after a vitrectomy procedure for recurrent retinal detachment, on post operative day 10 the surgeon found only 30% of gas was filled in patient¿s eye, they noticed the system was improperly dispensed gas and was leaking somewhere that caused room air to be drawn up.It was informed that an ophthalmic gas and an ophthalmic operating console were used in the procedure.The procedure was completed successfully.There was no patient harm reported.Additional information has been requested, but none received till date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN PERFLUOROPROPANE (C3F8) GAS
Common Device Name
INTRAOCULAR GAS
Manufacturer (Section D)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer (Section G)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15824177
Report Number2518435-2022-00036
Device Sequence Number6157583
Product Code LPO
UDI-Device Identifier00380657971053
UDI-Public00380657971053
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P900066
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065797105
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2022
Initial Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONSTELLATION VISION SYSTEM
Patient SexUnknown
-
-