| Model Number |
HYF-XP |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the customer, after a microbiological routine sampling of this device, carried out as required by french regulation, unexpected contamination was detected which was above the acceptable threshold for all channels.The test was performed prior to use.There was no contamination or any other deterioration in state of health of any person, to which this medical device could have been a contributory cause.Per customer hygiene microbiological investigation (hmi) sampling taken, the microbial results obtained are as follows: date of analysis (b)(6) 2022, sample taken (b)(6) 2022: mesophilic aerobic flora is greater than 80 cfu/volume for all channels.Declaration of conformity: the sample presented during its sampling characteristics that did not comply with the guides and recommendations of the ministry of health.Date of analysis (b)(6) 2022, sample taken (b)(6) 2022: mesophilic aerobic flora is 27 cfu/volume for all channels.Date of analysis (b)(6) 2022, sample taken (b)(6) 2022: mesophilic aerobic flora is 7 cfu/volume for all channels.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned but the evaluation is not yet begun.The device has been sent to an independent laboratory to perform a culture test for the device for olympus.The results are pending.Device return date is not available.Due diligence is being performed to collect more information from the customer.Supplemental report(s) will be submitted when any relevant new information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Device has been evaluated.This supplemental report is being submitted to provide this information.Please see the updates in sections: d9, g3, g6, h2, h3, h4, h6, and h10.Customer provided information on reprocessing performed at the facility: during pre-cleaning, suction is performed for all channels.Detergent used is salvanios.During manual cleaning detergent used is salvanios.Disinfectant used is anioxyde 1000.Brushing is performed for the instrument/suction channel, instrument channel port, and distal/areas around the elevator.Brushes used are of bytil novaclean brand.No information provided on machine reprocessing.The device is stored horizontally in a simple storage cabinet without drying feature.Maintenance of the device is by olympus.The device is not sterilized.An independent laboratory performed for olympus a culture test for the device after the device was reprocessed per the directions of the instructions for use.The test results for all channels indicate that the culture sampling results conform to the target levels established by the french regulation of july 4, 2016 with a number of revivable microorganisms less than 1 cfu/endoscope, for an endoscope subjected to high-level disinfection and rinsed with sterile water.Device history record review indicates that the device was manufactured and tested in accordance with all applicable procedures and met all final product release criteria at the time of shipping.The device is returned, and an evaluation completed for it.The user¿s complaint of contamination with microbes was not duplicated.There was no malfunction of the device that would contribute to the positive result of the microbial testing.It cannot be conclusively determined if the reprocessing performed by the customer contributed to the positive culture results.Upon inspection of the device it was observed that the device had light guide image bundle was broken.
|
| |
|
Search Alerts/Recalls
|
|
|