| Model Number |
000451 |
| Medical Device Problem Code |
Missing Information (4053)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the ellik evacuator was sterile but not marked anywhere on the outer wrap.Per additional information received on (b)(6)2022, it was reported that the ellik evacuator was leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the ellik evacuator was sterile but not marked anywhere on the outer wrap.Per additional information received on 15nov2022, it was reported that the ellik evacuator was leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿incorrect/ missing translation; missing instructions; vendor/printer error".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." corrections: g,h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
|
| |
|
Search Alerts/Recalls
|