• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EPIC VASCULAR; STENT, ILIAC Back to Search Results
Model Number 39200
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in the femoral artery of lower extremity.A 8x60x120 epic vascular stent was selected for use.However, during unpacking, the stent struts were found to be kinked.The procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the epic vascular 8x60x120 was returned for analysis.The device was received with the stent fully constrained and undamaged in the correct position on the delivery system.A visual examination identified no issues with the tip of the device.A visual and tactile examination found the shaft of the device to be severely kinked at approximately 240mm proximal of the distal tip.This type of damage is consistent with excessive force being applied to the device.A visual examination identified no issues or damage to the handle of the device.
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in the femoral artery of lower extremity.A 8x60x120 epic vascular stent was selected for use.However, during unpacking, the stent struts were found to be kinked.The procedure was completed with another of same device.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC VASCULAR
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15825272
Report Number2124215-2022-44918
Device Sequence Number17669445
Product Code NIO
UDI-Device Identifier08714729805687
UDI-Public08714729805687
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number39200
Device Catalogue Number39200
Device Lot Number0028203140
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/26/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexMale
-
-