• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TRANSPORT STRUT HOFFMANN LRF; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TRANSPORT STRUT HOFFMANN LRF; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 4933-0-400
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 08/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "patient came in and one of the struts was not transporting making the transporting bone shift and not line up to docking site.Patient had to spend an extra hour as the surgeon had replace the broken strut in clinic.This may delay his docking procedure in the future.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since doesn't matches the alleged failure, although the product was returned for evaluation.The optical and functional evaluation of the returned part shows that the part got returned in a strongly used, but functional condition.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: "patient came in and one of the struts was not transporting making the transporting bone shift and not line up to docking site.Patient had to spend an extra hour as the surgeon had replace the broken strut in clinic.This may delay his docking procedure in the future".
 
Additional Manufacturer Narrative
The complaint could be confirmed, since the product was returned for evaluation and matches the alleged failure.The optical and functional evaluation of the returned part shows that the part got returned in a strongly used, but functional condition.R&d has done first inspection and reviewed the received information and noted: there were samples which showed ¿deformations¿ which can be caused by one non-functioning strut in the frame construct.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a non-functioning strut in the construct.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSPORT STRUT HOFFMANN LRF
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15825314
Report Number0008031020-2022-00621
Device Sequence Number17697948
Product Code KTT
UDI-Device Identifier07613327094640
UDI-Public07613327094640
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K153377
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number4933-0-400
Device Catalogue Number49330400
Device Lot NumberD24256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer04/28/2023
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date05/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age35 YR
Patient SexMale
Patient Weight100 KG
-
-