| Catalog Number |
8065751762 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that before a cataract surgery system message was displayed after capsulorhexis was done.The surgery was postponed to next day.Surgery was completed after footpedal was borrowed.
|
| |
|
Additional Manufacturer Narrative
|
|
No service record relevant to the complaint reported event was found.There was no equipment returned for this investigation.Therefore, the reported issue cannot be confirmed.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate the manufacturer internal reference number is: (b)(4).
|
| |
|