Model Number 11104055
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The explanted device was not returned for evaluation but two radiographs were provided confirming the alleged device fault.The patients post op activity levels or whether they experienced a fall or other accident is unknown.Review of the radiographs provided identified the surgeon only used one pivoting rod to rod connector counter to instructions allowing the rod to toggle and disassociate from the connector which is considered to be the root cause.No additional investigation required.Related label review ".Potential adverse events and complications - as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries." ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation." ".Pivoting rod-rod connectors must be used in pairs (i.E., two per side)." ".Patient education: preoperative instructions to the patient are essential.The patient should be made aware of the limitations of the implant and potential risks of the surgery.The patient should be instructed to limit postoperative activity, as this will reduce the risk of bent, broken or loose implant components.The patient must be made aware that implant components may bend, break or loosen even though restrictions in activity are followed." ".Post-operative warnings -during the postoperative phase it is of particular importance that the physician keeps the patient well informed of all procedures and treatments.Damage to the weight-bearing structures can give rise to loosening of the components, dislocation and migration, as well as to other complications.To ensure the earliest possible detection of such catalysts of device dysfunction, the devices must be checked periodically postoperatively, using appropriate radiographic techniques." h3 other text : device not received.
 
Event or Problem Description
On (b)(6) 2022 a patient underwent a posterior fixation procedure on unknown levels of the spine.The surgery was completed without issue.On a unknow date it was discovered that a implanted rod detached from rotating connector.It is unknow if any adverse consequences were being experienced as a result of the device failure.On (b)(6) 2022 the patient underwent a revision procedure where the failed rotating connector was removed and replaced with an adjacent segment fixation connector.The patient is reported to be doing well post revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GSB GLOBAL SPINAL BALANCE SYSTEM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk blvd
san diego, CA 92121
MDR Report Key15825915
Report Number2031966-2022-00264
Device Sequence Number6143152
Product Code NKB
UDI-Device Identifier00887517596246
UDI-Public887517596246
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11104055
Device Lot NumberIA0582
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Initial Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
Patient Weight55 KG
-
-