C.R. BARD, INC. (BASD) -3006260740 BASIC KIT 18G X 10 CM WITH GUARDIVA; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Material Frayed (1262)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the customer "powerglide pro guidewire broke in patient.Nurses states they were placing a powerglide pro midline in a patient.When the nurse went to remove the guidewire, she felt like the tip of it got caught on the sheath.Once she got the guidewire out, nurse states it looks like the tip has broken off.Nurse is asking what type of imaging to perform to see if part of the guidewire is still in the patient.Informed nurse the guidewire is made of nitinol.It will show up on x-ray, however they should consider the possibility of it migrating when they scan it.Informed nurse she should discuss with the provider the material of the guidewire as which imagining to order would be a clinical decision the provider should make.".
 
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of refs2813 showed no other similar product complaint(s) from this lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASIC KIT 18G X 10 CM WITH GUARDIVA
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15825967
Report Number3006260740-2022-05373
Device Sequence Number6156865
Product Code FOZ
UDI-Device Identifier00801741140464
UDI-Public(01)00801741140464
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K162377
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model NumberN/A
Device Catalogue NumberF118107T
Device Lot NumberREFS2813
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/07/2022
Initial Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-