The device was returned.Based on investigation results, it is determined that liquid could not be injected due to compressive buckling of the needle tube.Since friction resistance between the outer tube and needle tube has increased, compressive buckling might have occurred when the needle was extended.A likely mechanism causing friction resistance between outer tube and needle tube might be the following: the tube was coiled during the inspection for operation.The root cause of this event was unable to be identified.A review of the device history record found no deviations that could have caused or contributed to the reported issue.No complaint trend abnormalities were detected in the quality data analysis of the event.The event can be prevented by following the instructions for use which state: ·straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Operate the slider slowly, otherwise the tube could buckle.Do not coil the insertion portion with a diameter of less than 15 cm.This could damage the insertion portion.When inserting the instrument into the endoscope, retract the needle into the sheath, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.·insert the instrument slowly.Abrupt insertion could damage the endoscope and/or instrument.·stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Olympus will continue to monitor field performance for this device.
|