| Model Number |
C503 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable results from the cobas pro c 503 analytical unit.Patient 1 initial bild2 (direct bilirubin) result was 0.714 mg/dl and the repeat result was 0.0922 mg/dl.Patient 2 initial gluc3 (glucose) result was 5.72 mg/dl and the repeat result was 85.5 mg/dl.The questionable results were not reported outside of the laboratory.The direct bilirubin reagent lot number was 63837501 with an expiration date of 31-aug-2023.The glucose reagent lot number was 61404201 with an expiration date of 31-may-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
The calibration and qc data were acceptable, therefore a general reagent issue could be excluded.The alarm trace showed several abnormal aspiration and clot alarms.Therefore, the issue was consistent with a hardware/maintenance issue, and also with a sample quality (preanalytic) issue.
|
| |
|