| Model Number |
989605384711 |
| Medical Device Problem Code |
Device Tipped Over (2589)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The philips service engineer determined that the system tipped over when encountering the barrier between the elevator and the floor during transport.This issue was not caused by a malfunction of the system.The users were reminded to transport the system with caution.
|
| |
|
Event or Problem Description
|
|
A customer reported their cx50 ultrasound system tipped over after encountering an obstacle during transport within the user facility.There was no allegation of patient or user injury as a result of the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
An addendum report is being submitted to provide the date received by manufacturer with the philips¿ become aware date.The information is in the g3 field.
|
| |
|