| Model Number |
F14NC |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Urinary Tract Infection (2120)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Common device name: catheter, straight.Based on the available information, this event is deemed to be a serious injury.This complaint has been evaluated.No lot number is available.A detailed investigation or batch review cannot be conducted.Therefore, this evaluation will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: (b)(4).Manufacturing site: (b)(4).
|
| |
|
Event or Problem Description
|
|
End user reports recurrent utis for last 3 months with use of f14nc catheters.She said "she really isn't sure why she has been getting so many utis all of a sudden these last 3 months.She has been using these catheters for many years during her 15 years cathing and didn't get utis prior with use of these.She caths 6 x a day following nerve damage to her bladder following a tumor that was removed from her spine.She said she did have kidney stones during these last 3 months she got surgically taken care of and she thought that was to blame for her utis as she knows stones will place her at risk for utis.She said she has to follow up with her urologist to follow up post surgically as she was recently x-rayed to ensure all stones gone.She said she hasn't had typical uti symptoms and denies all other uti symptoms with her only symptom is having a "draggy feeling" and just doesn't "feel well" so she calls her urologist and they have her do a urine sample and they have her treated with antibiotics based on urine results showing utis.She mentioned she has had amoxicillin, bactrim rounds during last 3 months she has had and she is currently taking a treatment round of bactrim since last thursday for her lastest uti.She said she goes for urine testing after her antibiotic rounds and she said her urine looks like there is no uti per her urologist then 1 or 2 weeks goes by and she gets another one.She said this most recent uti she is being treated for was right after she went home from the hospital as she recently had shoulder surgery.She said she was tested prior to going into the hospital and she had no uti.When asked if they had placed a foley catheter in her for the surgery she said she didn't remember one.She thought "maybe the material has got something to do with it" clarified with her possible concern of the material's sterility.She confirmed with me they were all packaged with intact packaging prior to use.Nothing abnormal about these catheters.No pain or discomfort with use as well.She washes her hands prior to cathing, she makes sure catheter stays sterile prior to insertion and only holds the very end of it.She said she also wipes her urethra with an antibacterial wipe prior to insertion.She never has used lubricant with these catheters stating she "was never instructed to." end user was advised her we do advise to use lube with uncoated catheters." this emdr covers the unknown number of market units and unknown number of catheters associated with the utis.
|
| |
|
Search Alerts/Recalls
|