| Model Number |
A2000 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility reported that the mayfield 2000 skull clamp (a2000) was on a patient and there was a little movement of the locking mechanism.There was no patient injury noted and no delay in surgery has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The mayfield skull clamp (a2000) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - the complaint is confirmed via inspection of the unit.The investigation showed that the lock had rotational and lateral movement due to a worn hex bushing.To resolve the issue, new components were added to replace all worn internal parts, and general maintenance and cleaning has been performed.Root cause - root cause is routine use and wear of the unit as the bushing was worn resulting in movement in the lock while in the locked position.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|
| |
|