| Model Number |
Fibula Cutting Guide |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the design files for the fibula cutting guide found that the guides were designed correctly to the specifications and the surgical plan performed by the surgeon.
|
| |
|
Event or Problem Description
|
|
During a reconstruction surgery featuring virtual surgical planning, the use of a fibula cutting guide led to a fibula segment which was reported to be too large for the mandible by approximately 5 mm.The surgeon manually cut the segment again to achieve proper fit and reported a 60-minute delay in the procedure.
|
| |
|