• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR Back to Search Results
Model Number PROPAQ MD
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for defib and pacer functions.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including full functional testing, bench handling and defib cycling without duplicating the report.The processor/bridge/pace board was replaced and scrapped as a precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD DEFIBRILLATOR
Common Device Name
PROPAQ MD
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key15828339
Report Number1220908-2022-04452
Device Sequence Number17681030
Product Code MKJ
UDI-Device Identifier00847946017699
UDI-Public00847946017699
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P160022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPROPAQ MD
Device Catalogue NumberPROPAQ MD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer10/31/2022
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date01/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-