TELEFLEX MEDICAL VLOCK ML CLIP 6/CART 14/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN055290
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Intra-Abdominal Hemorrhage (4479)
Date of Event 10/29/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
Reported issue: lap chole went normal, no issues occluding vessels and dividing.Post op patient came back with blood in the abdomen.Surgeon coud not be 100% sure it was the cystic artery but believes that is where the bleeding came from the clip breaking/falling off.Believes the vesolok is more brittle and does not have the same tactile closure as hemolok.The patient's condition was reported as fine.
 
Additional Manufacturer Narrative
(b)(4).The customer returned one unopened representative sample of 51114v vlock ml clip 6/cart 14/box for investigation.No broken clip or clip applier was returned.The returned sample was visually examined with and without magnification.No defects or anomalies were observed.Functional inspection was performed on the returned clips.A lab inventory clip applier was used.All clips were able to properly load into the jaws of the applier and were successfully applied to over-stressed surgical tubing without detaching.It is possible that the reported issue was caused by using damaged/misaligned appliers, but this could not be confirmed since the appliers were not returned for investigation.No functional issues were found with the returned clips.The finished goods dhr was reviewed.Per the dhr, lot number: 228044 of product 51114v vlock ml clip 6/cart 14/box was manufactured on 16mar2022.A total of (b)(4) units were manufactured.The lot was released on 08apr2022.Dhr investigation did not show issues related to complaint.The ifu for this product, 15-lb-00287 rev.00, was reviewed as a part of this complaint investigation.The ifu states, "manual-load applier/remover jaws are delicate and can easily become damaged, as can some other applier/remover components.Mishandling of applier/removers may result in improper load and/or closure of the jaws.Appropriate care, cleaning, lubrication and maintenance are important to ensure proper function.Please examine the applier/remover before each surgery for potential damage." a corrective action is not required at this time as there were no functional issues found with the returned representative sample.The reported complaint of "broken/detached parts - clip - hinge" could not be confirmed.One representative cartridge was returned.Upon functional inspection, the clips could be loaded into the jaws of a lab inventory applier and were able to properly close onto over-stressed surgical tubing with no issues.A dhr review was performed with no evidence to suggest a manufacturing related issue.Since there were no functional issues with the returned clips, the complaint cannot be confirmed.Teleflex will continue to monitor and trend on complaints of this nature.
 
Event or Problem Description
Reported issue: lap chole went normal, no issues occluding vessels and dividing.Post op patient came back with blood in the abdomen.Surgeon could not be 100% sure it was the cystic artery but believes that is where the bleeding came from the clip breaking/falling off.Believes the vesolok is more brittle and does not have the same tactile closure as hemolok.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VLOCK ML CLIP 6/CART 14/BOX
Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
9194332672
MDR Report Key15828510
Report Number3011137372-2022-00219
Device Sequence Number12301283
Product Code FZP
UDI-Device Identifier24026704666823
UDI-Public24026704666823
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2025
Device Model NumberIPN055290
Device Catalogue Number51114V
Device Lot Number228044
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age29 YR
Patient SexFemale
Patient Weight103 KG
-
-