3M HEALTH CARE 3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW; HIGH FLOW DISPOSABLE TUBING Back to Search Results
Catalog Number 24355
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of events are unknown.3m has investigated and identified a manufacturing issue with the auto-venting bubble trap that introduces the risk of blood or fluid leaks while priming the sets and/or during fluid administration.There is no alleged injury to the patient or caregiver for this reported incident.3m has initiated a field safety corrective action for lots associated with this issue.3m has implemented a change to address the identified manufacturing issue and will continue to monitor.
 
Event or Problem Description
A hospital alleged 3m¿ ranger¿ high flow warming set (model 24355) lot hx9190 leaked at the bubble trap on multiple occasions.It was indicated that a pressure device was used.No patient or staff injury was alleged.No medical interventions were required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3M¿ RANGER¿ FLUID WARMING SET, HIGH FLOW
Common Device Name
HIGH FLOW DISPOSABLE TUBING
Manufacturer (Section D)
3M HEALTH CARE
2510 conway ave
st. paul MN 55144
Manufacturer (Section G)
NYPRO PRECISION ASSEMBLIES
sor juana ines de la cruz
cd. industrial
tijuana, tijuana baja california 22444
MX   22444
Manufacturer Contact
dianne gibbs
3m center, building 275-5w-06
2510 conway ave
st. paul, MN 55144
6517379117
MDR Report Key15828717
Report Number2110898-2022-00116
Device Sequence Number14988354
Product Code BSB
UDI-Device Identifier00707387781446
UDI-Public00707387781446
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K973741
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/25/2025
Device Catalogue Number24355
Device Lot NumberHX9190
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2022
Initial Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-