MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM NRHFEM W/PIN MED RT 71X66; LPS AND S-ROM : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM NRHFEM W/PIN MED RT 71X66; LPS AND S-ROM : KNEE FEMORAL Back to Search Results
Model Number 62-3401R
Medical Device Problem Codes Loss of or Failure to Bond (1068); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Unspecified Infection (1930); Pain (1994); Joint Laxity (4526)
Date of Event 11/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision completed due to patient falling on his knee.Upon exposure to the joint space it was discovered that medium srom hinged femur had loosened away from the porous femoral sleeve construct.The loosening occurred at the taper point of the femur to the sleeve.Surgeon removed the still engaged hinged pin with a pair of kochers and then removed the medium x 26mm poly with his hand and then removed the medium right femur by hand.Upon further examination the surgeon stated that none of the implants had broken or shown unusual wear but had simply disengaged as a result of the fall the patient experience.The tibial component and femoral sleeve/stem were well fixed.Doi: unknown; dor: (b)(6) 2022; affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (nc search) was performed for the finished device 623401r, lot - j5925n, and no non-conformances / manufacturing irregularities were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM NRHFEM W/PIN MED RT 71X66
Common Device Name
LPS AND S-ROM : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15829291
Report Number1818910-2022-23286
Device Sequence Number6155364
Product Code KRO
UDI-Device Identifier10603295179474
UDI-Public10603295179474
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K896048
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number62-3401R
Device Catalogue Number623401R
Device Lot NumberJ5925N
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS UNIV TIB HIN INS MED 26MM; UNKNOWN KNEE FEMORAL SLEEVE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-