| Model Number |
L-370 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device serial number/lot number is unknown, no product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer was having issues with the connection not being completely connected.The unit was leaking fluid due to the new tubing.This has been happening often with the new system.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: b5: additional information.H6: event problem and evaluation codes: updates not required.H10: device evaluation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
Additional information received on 09-dec-2022 via email and attached to complaint object: the customer reported that we need to disregard this issue.They found from the technical department that they need to make sure to push in the line all the way and this was very difficult to do.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: event problem and evaluation codes: updates not required.H10: device evaluation was completed.No product was returned for investigation.The cause of the reported problem could not be determined.No lot number was provided; therefore, dhr (device history review) could not be performed.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 if additional reportable information becomes available.No lot or serial numbers were communicated; d4: device expiration date and h4: device manufacturing date are not available.
|
| |
|
Additional Manufacturer Narrative
|
|
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Search Alerts/Recalls
|