| Model Number |
TPS-101-1112 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that the ps femoral implant packaging had a hole through both layers of packaging, breaching the sterile barrier.This was discovered during surgery; the surgeon chose to re-sterilize the implant.This re-sterilization process takes around 1-2 hours and needed to be completed while the patient was under anesthesia.This was a major delay to the surgery.The device history record was reviewed, and the implant was designed and manufactured per the specifications.All sterilisation requirements were met.
|
| |
|
Event or Problem Description
|
|
It was reported that the ps femoral implant packaging had a hole through both layers of packaging, breaching the sterile barrier.This was discovered during surgery; the surgeon chose to re-sterilize the implant.This re-sterilization process takes around 1-2 hours and needed to be completed while the patient was under anesthesia.This was a major delay to the surgery.
|
| |
|