Model Number 750180
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that the cuff could not be fully expanded, problem discovered during testing upon opening.No injury reported.
 
Additional Manufacturer Narrative
H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample and five photos were received for evaluation.Functional testing found the cuff was fully inflated and symmetrically; complaint was not confirmed.Device passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TUBES ADULT AIRE-CUF
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
NULL
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15829424
Report Number3012307300-2022-27338
Device Sequence Number5100570
Product Code JOH
UDI-Device Identifier15021312006421
UDI-Public15021312006421
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2024
Device Model Number750180
Device Catalogue Number750180
Device Lot Number3748041
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2022
Supplement Date Received by Manufacturer01/26/2023
Initial Report FDA Received Date11/18/2022
Supplement Report FDA Received Date05/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
Patient RaceAsian
-
-