| Catalog Number |
5955600 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As reported, during a robotic hernia repair the yellow anchor of the ventralight st echo ps, got stuck in the patient's fascia when pulling the inflation tube through abdominal wall.Often during robotic repairs, the primary defect is closed prior to placing the mesh, this can inhibit the passage of the inflation tube while pulling through the anatomy.Note it is unknown if the primary defect was closed in this case.Evaluation of the returned echo ps assembly found visible material deformation at the weld site on the inflation tube.The deformation found is from the rf weld performed on the anchor and serves as evidence that the anchor / weld process was properly performed.Based on the sample evaluation and investigation performed, the most probable root cause is that resistance was encountered when pulling the inflation assembly through the patient¿s anatomy causing the detachment at the anchor.To date, this is the only reported complaint for this manufacturing lot of 116 units released for distribution in june 2022.A review of manufacturing records was performed and confirms product was made to specification.The instructions-for-use (ifu) supplied with the device states: "once the ventralight st mesh w/echo ps positioning system (the device) is inserted into the abdomen, use a grasper to locate the blue retrieval loop on the inflation tube, ensuring that the blue inflation tube is not wrapped around the mesh and is clearly visible.Pass a suture passer device through healthy skin at the center of the hernia defect (avoid going directly through the umbilicus).Grasp the blue retrieval loop and pull the retrieval loop and inflation tube out of the abdominal cavity.Sample evaluated.
|
| |
|
Event or Problem Description
|
|
As reported, during a robotic umbilical hernia repair on (b)(6) 2022, a ventralight st mesh w/echo ps was implanted through 12mm trocar to the patient.It is reported that after insertion, when the surgeon was pulling the inflation tube up through the abdominal wall skin incision, the inflation tube broke, and the yellow anchor got stuck in the patient's fascia.As reported, the surgeon was able to inflate the echo ps and fixate the mesh without any issues; however, after they removed the echo ps once the mesh was fixated, they noticed the yellow anchor was missing so the surgeon made an extension of the skin incision to retrieve the yellow anchor and piece of inflation tube.No patient harm or injury was reported.
|
| |
|
Search Alerts/Recalls
|
|
|